產(chǎn)品
2B臭氧分析儀202型使用254nm紫外光吸收法,適用于監(jiān)測(cè)大氣中的臭氧濃度,量程由1.5ppbv至100ppmv。分析儀特點(diǎn)高精度、體積細(xì)小、重量(4.7磅,2.1公升)及低電量消耗。備有遙控監(jiān)測(cè)配件。
特點(diǎn):
使用254nm紫外光吸收法
高精度及分解度,量程由1.5ppbv至100ppmv
低電量消耗,能以12VDC(5.0W)外置電源操作
DewLineTM技術(shù)減低水氣對(duì)儀器的干擾
模擬輸出0-2.5V
應(yīng)用:
氣球及航空的大氣測(cè)量
建立城市監(jiān)測(cè)網(wǎng)絡(luò)
個(gè)人及工作環(huán)境監(jiān)測(cè)
工業(yè)生產(chǎn)中監(jiān)測(cè)
產(chǎn)品規(guī)格:
操作原理
紫外光254nm燈
量程
1.5ppbv-100ppmv
精度
2%
采樣時(shí)間
10秒、1分鐘、5分鐘及1小時(shí)
記憶容量
20,480組
輸出
數(shù)碼顯示,0-2.5V,RS-232
流量
大約1L/min
尺寸/重量
90 210 290mm/4.7lb
電源
12VDC交流電
The2BTechnologiesModel202OzoneMonitor isdesignedtoenableaccurateandprecisemeasurementsofozonerangingfromlowppb(precisionof~1ppb)upto100,000ppb(0-100ppm)basedonthewellestablishedtechniqueofabsorptionofUVlightat254nm. TheModel202OzoneMonitor islightweight(4.7lb.,2.1kg.)andhasalowpowerconsumption(12VDC,0.33amp,4.0Watt)relativetoconventionalinstrumentsandisthereforewellsuitedforapplicationssuchas:
Verticalprofilingusingballoons,kites,RPVsandlightaircraftwherespaceandweightarehighlylimited
Long-termmonitoringatremotelocationswherepowerishighlylimited
Urbanarraysofground-baseddetectors
Personalexposuremonitoringforstudiesofhealtheffectsofairpollutants
Environmentalhealthandsafetymonitoring
Laboratorystudiesoftheeffectsofozoneexposureonmaterialsandorganisms
TheModel202hasbeendesignatedbytheEPAasaFederalEquivalentMethod(FEM):EQOA-0410-190.
Options
FlashCardMemoryforvirtuallyunlimited,portabledatastorage
GPSforcontinuousloggingoflatitude,longitudeandaltitude
Rechargeablemetalhydridebatteryprovides12hrofcontinuousoperation
Externalparticlefilter
Serial-to-USBadapter
Rotaryvaneairpumpforlowtemperatureandquietoperation
Externalairpumpforlongpumplife
Specifications
MeasurementPrinciple
UVAbsorptionat254nm
LinearDynamicRange
1.5ppbto250ppm
Resolution
0.1ppb
Precision(1 ;rmsnoise)
Greaterof1.5ppbor2%ofreading
Accuracy
Greaterof1.5ppbor2%ofreading
LimitofDetection(2 )
3.0ppb
NIST-TraceableCalibration
Yes
MeasurementInterval
10s(Dataaveragingoptions:10s,1min,5min,1hr)
FlowRate(nominal)
~1Liter/min
FlowRateRequirement
0.6L/min
BaselineDrift
2ppb/day
5ppb/year
SensitivityDrift
1%/day
3%/year
MeasurementTime,Frequency
10s,0.1Hz
ResponseTime,100%ofStepChange
20s,2points
AveragingTimes
10s,1min,5min,1hr
DataStorage
14,336lines(10savg.=1.4days;5minavg=1.4mo.)
OzoneUnits
ppb,ppm,pphm, gm-3,mgm-3
PressureUnits
mbar,torr
TemperatureUnits
C,K
TandPCorrected
Yes
OperatingTemperatureRange
0to50 C;-20to50 Cwithlowtemperaturemodifications(rotaryvanepumpandlampheater)
OperatingAltitudeRange
~0-30km(~30-1000mbar)
PowerRequirement;
Suppliedbybatteryor110/220VACPowerPack
12Vdcor120/240Vac,335mA,4.0watt(2.9wattwithcellheaterunplugged)
Size
3.5x8.5x11inches(9x21x29cm)
Weight
4.7lb(2.1kg);1.6lb(0.7kg)withoutinstrumentcase
DataLoggerCapacity
14,336lines(10savg.=1.4days;1minavg=10days;5minavg=1.4mo;
1hravg=1.6yr
AnalogInputsforInternalLoggingofOtherInstruments
3AnalogInputs,0-2.5V;ForexamplecouldlogexternalT,PandRH
DataOutputs
RS232,0-2.5VAnalog,LCDDisplay;ExternalUSBConverter:(Optional4-20mACurrent;requestquote)
DataTransferBaudRates
1200,2400,4800,19200
OutputRanges
UserDefinedScalingFactorinMenu
DewLine
Yes
BackupAirPump
Yes
FlowMeter
Yes
Options
GPS,FlashCardMemory,4-20mACurrentOutput;LampHeater;RotaryVaneAirPump;Long-LifeExternalAirPump;RackMountCase
UpgradabletoModel205DualBeam
Yes
SystemIncludes
Model202OzoneMonitor
ACPowerAdapter(100-240VACto12VDC) withCountry-SpecificPlug
CigaretteLighterAdapter
BatteryWireConnector
ZeroingCartridge
OperationManualonCD
CalibrationDataandNIST-TraceableCalibrationCertificate
InstrumentBirthCertificate
OneYearWarranty
TodownloadabrochuredescribingtheModel202OzoneMonitor (PDFformat),clickonthedownloadbuttonbelow.